First, the flowmeter without display flow solution
(1) First check whether there is a problem with the line, such as the signal line falling off or broken.
(2) Separate the sensor from the signal amplifier, connect the signal amplifier to the instrument, and use iron metal to move back and forth at a distance of 2 ~ 3mm from the bottom of the amplifier to take the signal. Please remove the flow sensor from the pipeline and check whether the flowmeter impeller is entangled or the impeller is damaged.
Second, the flow fluctuation is relatively large
The reasons for the relatively large flow fluctuations can be considered from the following aspects:
1. Whether there is interference
There are interference sources such as motors, frequency converters, and strong electricity near the flowmeter. Solution: ground the flowmeter or add filter capacitor. If it cannot be solved, it should be far away from the interference source.
2. Whether the water source is pulsating flow
(1) Generally speaking, the pump is used to pump raw materials. If the flow installation position is relatively close to the pump, it is easy to produce pulsating flow, which causes large fluctuations in the flow rate. Stabilize the flow.
(2) The installation position of the flowmeter is too close to the position of the valve or elbow. When the raw material passes through the valve or elbow, the flow rate fluctuates. At this time, it should be far away from the position of the valve and the elbow to ensure a certain length of straight pipe before and after.
Three, the flowmeter display error is relatively large
(1) First check whether the flow sensor coefficient, namely K value, and Other parameters of the instrument are set correctly
(2) When the customer has conditions, use an electronic scale for actual calibration.
Fourth, the flowmeter shows a small flow, but the actual flow is relatively large
The general cause of this problem is that the impeller does not rotate fast or the blade breaks. Remove the flow from the pipeline and check whether the flowmeter is entangled or damaged.